Vacuum Freeze Dryer, Lyophilizer Common Breakdown & Elimination

Jun 03, 2016 Leave a message


Vacuum freeze dryer is also popular as lyophilizer, its common breakdown and elimination as below:


1.The vacuum can not achieve below 15Pa:

1) Check the release valve is whether fastened down or not.

2) The vacuum pump and main engine is connected by vacuum tube. Check the clamp is whether fastened down or not.

3) Check the bottom of organic glass cover is whether smooth or not. Check the “O” rubber ring is whether damaged or not.

4) Check the vacuum pump oil. Generally, oil should be changed after working 200 hours.


2.The vacuum pump of vacuum freeze dryer leaks oil:

Check the leaking part (is possible sealed pad). Replace the corresponding parts.


3.The temperature of cold trap is high:

The temperature of condition is high. If the cooing system is break down, please connect with our company engineer.

More vacuum freeze dryer customization welcome contact salesj@toptionlab.com 24h/7days service.