Thermostatic Shaker, How To Clear Simple Faults Of Thermostatic Shaker? Lab Supplies
TOPTION thermostatic shaker is widely applied to the fields which have high requirements of temperature and oscillation frequency, such as biological, chemical reaction, bacteria culture, fermentation, enzyme, cell and tissue research, etc.
For microbial cells and various strains of exercise training, can also be used for static cultivation, in biology, molecular, such as medical, pharmaceutical, food, environmental protection has extensive and important application research.
During your normal thermostatic shaker operation, maybe will meet some simple faults, so how could we remove these simple fault?
Faults | Reason | Elimination method |
Instrument is not running, panel no display | Power supply not connect | Instrument power supply whether plugged in, Whether the power socket with electricity, The power switch whether is open. |
The fuse burned | Change same size fuse | |
Instrument running, LCD screen no display; LCD screen no display | In the same line has high frequency interference sources. | Click on the parameter correct/confirm button, it will be display again; Eliminate same line interference sources or use the other dedicated line. |
Large temperature fluctuations | Refrigerating method setting is incorrect. | According to the manual instruction reset. |
The measured temperature is high and often into the overtemperature alarm. status | Start line to control the temperature is too high. | Open the instrument back side aspirail or start refrigeration key. |
Large instrument noise. | Instruments placed not flat. | Adjusting instruments' left foot, make instrument be smooth landing. |
Porous tray is loose. | Tighten the fastening nut which is at the four corners of tray. | |
Flask clip loose | Take out the tray, with attached random tool fastening nut of flask. |